Introduction
CLS Users Guide 11
Miscellaneous Specifications
Serial Interface
Type:
RS-232 3 wire or RS-485 4 wire.
Isolation:
RS-232 None
RS-485 To EIA RS-485 Specification.
Baud Rate:
2400 or 9600, user selectable.
Error Check:
BCC or CRC, user selectable.
Number of Controllers:
1 with RS-232 communications;
32 with RS-485 communications.
Protocol:
Form of ANSI X3.28-1976 (D1, F1), compatible with Allen
Bradley PLC, full duplex.
System Power Requirements
Voltage:
12-24 Vdc
Input Current (no load):
300 mA max
Maximum Current Requirement:
610 mA. (If the reference voltage is
externally loaded, add 1 mA supply current for every 1 mA of load up to
a maximum load of 100 mA. If using the +5V logic supply to power
digital outputs, add 0.6 mA supply current for every 1 mA of load up to
a maximum load of 350 mA. Therefore, the maximum current
requirement is 300 +100 + (0.6 x 350) = 610 mA.)
Environmental Specifications
Storage Temperature:
-20 to 60
°
C
Operating Temperature:
0 to 50
°
C
Humidity:
10 to 95% non-condensing.
Physical Dimensions
CLS:
1.75 lbs., 1.98
"
x 3.78
"
x 7.10
"
(.8 kg, 50 mm x 96 mm x 180
mm.)
TB-18:
1.025
"
x
3.700
"
(2.57 cm x 9.29 cm)
TB-50:
3.2
"
x 3.4
"
(8.03 cm x 8.53 cm)